Product picture SS2100a TDLAS gas analyzer, right angle view

SS2100A

SS2100A
Please Wait (spinning wheel)
Fiyat yükleniyor
Fiyatlar geçici olarak mevcut değil
Price on request
fiyat etkinleştiriliyor
fiyat: bilgi verin
Ölçüm görevini/uygulama görevini filtrele
Ölçüm / Uygulama görevi

Bu ürünün CAD çizimlerini indirmek için yapılandır

  • Measuring principle

    TDLAS

  • Channels

    1

  • Analyte and Measurement ranges

    H2S (Hydrogen Sulfide):
    0 to 10
    0 to 1000 ppmv
    0 to 5000 ppmv
    0 to 5% volume

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz; 60W max (with 2 solenoids) - electronics enclosure
    120 or 240 VAC, 50 to 60 Hz - standard; 100W or 200W max for heated systems - sample cabinet

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
    Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: Copper-free Aluminum
    Sample System Enclosure: 304 or 316 stainless steel

  • Hazardous area approvals

    ATEX Zone 2

  • Degree of protection

    IP66

  • Product safety

    CE

  • Measuring principle

    TDLAS

  • Channels

    1

  • Analyte and Measurement ranges

    H2O (Moisture):
    0 to 10 ppmv
    0 to 100 ppmv
    0 to 50 ppmv
    0 to 5000 ppmv

  • Measured Variables

    Concentration
    Moisture Dew Point
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz; 60W max (with 2 solenoids) - electronics enclosure
    120 or 240 VAC, 50 to 60 Hz - standard; 100W or 200W max for heated systems - sample cabinet

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
    Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: Copper-free Aluminum
    Sample System Enclosure: 304 or 316 stainless steel

  • Hazardous area approvals

    ATEX Zone 2

  • Degree of protection

    IP66

  • Product safety

    CE

  • Measuring principle

    TDLAS

  • Channels

    1

  • Analyte and Measurement ranges

    CO2 (Carbon Dioxide):
    0 to 10 ppmv
    0 to 1000 ppmv
    0 to 5000 ppmv
    0 to 5% volume
    0 to 10% volume
    0 to 20% volume

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz; 60W max (with 2 solenoids) - electronics enclosure
    120 or 240 VAC, 50 to 60 Hz - standard; 100W or 200W max for heated systems - sample cabinet

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
    Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: Copper-free Aluminum
    Sample System Enclosure: 304 or 316 stainless steel

  • Hazardous area approvals

    ATEX Zone 2

  • Degree of protection

    IP66

  • Product safety

    CE